Philips IE33 ultrasonic diagnostic instrument common fault troubleshooting(6)


Fault phenomenon
The probe is not fully connected to the device, resulting in loss of crystal oscillator signal or other image quality problems.

Troubleshooting
The inner hexagon screw used for internal fixation of probe interface has slipped and the probe cannot be fully connected with the probe interface module on the machine.Use inch 3/32 socket head tool with set screws.