Philips IE33 ultrasonic diagnostic instrument common fault troubleshooting(2)
 
Fault phenomenon
System error"602"

Troubleshooting
Most of the 602 errors are due to memory: loose connections between memory and slots, or poor contact due to oxidation or dirt on the surface of memory gold fingers.First eliminate 602 errors due to incorrect BIOS memory hole setup (use FII Instruction to verify the hard-ware /Software memory hole setup for the current Software version).After cleaning and refixing the memory, the equipment started normally, troubleshooting.