Medical News

Philips IE33 ultrasonic diagnostic instrument common fault troubleshooting

Views : 436
Update time : 2020-08-27 17:33:52

Philips IE33 ultrasonic diagnostic instrument common fault troubleshooting
 

 

Fault phenomenon

The device failed to start, stopped at the "initializing hardware" screen, or error 606 was reported during startup.
 

Troubleshooting

The motherboard battery is dead and the BIOS Settings will automatically jump to the Default Settings, which will cause an exception to the HOST or SIP boot.Mainboard battery life is about 5 years, use a multimeter to measure the battery voltage, less than +2.85Vdc, replace the mainboard battery (HOST/SIP mainboard battery model is CR2032, UMB/EMB mainboard battery model is CR2450), and change the BIOS to Philips Settings, troubleshooting.

Related News
Read More >>
Hitachi Aloka HI VISION PREIRUS: Image Interference Hitachi Aloka HI VISION PREIRUS: Image Interference
Jan .31.2024
Image Interference Troubleshooting for Hitachi Aloka HI VISION Preirus
GE VIVID I system maintainance GE VIVID I system maintainance
Jan .30.2024
Ultrasound system,Hospital equipment,Medical diagnosis implement,Imagine parts,Medical spare parts,Ultrasonic Sensor,Hitachi Aloka,GE,Siemens,Toshiba,Philips
About Us
As one of the leading ultrasound service companies in China, Rongtao Medical provides healthcare solutions, repair service, parts and accessories to hospitals, clinics and distributors world-widely.
Connect with us

Copyright 2014-2023 Guangzhou Rongtao Medical Tech LTD All Rights Reserved.