Medical News

Philips CX50 report error 0022, can not enter the system

Views : 1388
Update time : 2023-02-27 16:25:00

Failure phenomenon:

Philips CX50 report error 0022, can not enter the system.

 

Failure analysis:

When the machine starts up, error 0022 is reported, which may be caused by the failure of the power board or channel board. First enter the background self-check, and the self-check result is the problem of the Pwr Sup power board, so the next troubleshooting of the power board.

 

Repair results:
Replace the power supply, all the functions of the machine are used normally, and the fault is eliminated.

Related News
Read More >>
Hitachi Aloka HI VISION PREIRUS: Image Interference Hitachi Aloka HI VISION PREIRUS: Image Interference
Jan .31.2024
Image Interference Troubleshooting for Hitachi Aloka HI VISION Preirus
GE VIVID I system maintainance GE VIVID I system maintainance
Jan .30.2024
Ultrasound system,Hospital equipment,Medical diagnosis implement,Imagine parts,Medical spare parts,Ultrasonic Sensor,Hitachi Aloka,GE,Siemens,Toshiba,Philips
About Us
As one of the leading ultrasound service companies in China, Rongtao Medical provides healthcare solutions, repair service, parts and accessories to hospitals, clinics and distributors world-widely.
Connect with us

Copyright 2014-2023 Guangzhou Rongtao Medical Tech LTD All Rights Reserved.