Medical News

GE LOGIQ E9: Frequent Crashes

Views : 473
Update time : 2020-10-16 11:07:22
 
Failure phenomenon
GE LOGIQ E9 color Doppler ultrasound frequently crashes after turning on.

Failure analysis
This fault is a startup fault. First, it is necessary to determine whether the fault is at the front or back end of the system.

Open the BEP shell, turn it on, and find that the rear power indicator shows the following: PS-ON and 5 V stdby LED indicators are on, other indicators are off. It can be seen that +5 V STB has output and PS-ON has action, but there is no +48 V voltage directly output by the main power supply. The reasons for this phenomenon may be:
① The main power supply fails;
② The +48 V load has an overcurrent short-circuit phenomenon.

After disconnecting the +48 V load, the fault remains. It is judged that the fault is caused by poor main power supply. For further confirmation, after the main power supply was removed and tested separately, it was found that the main power supply was damaged and there was no +48 V output. After replacement, the fault was eliminated.
Related News
Read More >>
Hitachi Aloka HI VISION PREIRUS: Image Interference Hitachi Aloka HI VISION PREIRUS: Image Interference
Jan .31.2024
Image Interference Troubleshooting for Hitachi Aloka HI VISION Preirus
GE VIVID I system maintainance GE VIVID I system maintainance
Jan .30.2024
Ultrasound system,Hospital equipment,Medical diagnosis implement,Imagine parts,Medical spare parts,Ultrasonic Sensor,Hitachi Aloka,GE,Siemens,Toshiba,Philips
About Us
As one of the leading ultrasound service companies in China, Rongtao Medical provides healthcare solutions, repair service, parts and accessories to hospitals, clinics and distributors world-widely.
Connect with us

Copyright 2014-2023 Guangzhou Rongtao Medical Tech LTD All Rights Reserved.